testo 335 compact flue gas analyser
Stock Code:
UCT-FH06323350
Equivalent to: 
testo 06323350

Enquire

Description

testo 335 is the new generation flue gas analyser, specially tailored to industrial applications' requirements. testo 335 can be used for all emission monitoring applications by the operators of industrial furnaces such as processing and power plants, by service technicians for burner/furnace manufacturers, for plant construction as well as for stationary motors. Even spot measurements for up to 2 hours are possible.

Feature:

  • max. 3 measurement cells
  • Two toxic measurement cells, freely selectable
  • High-performance automatically controlled membrane pump
  • Logger function
  • Parallel ΔP and m/s for flue gas measurement
  • 2 measurement range extensions (standard: single dilution with factor 5, option:dilution for all sensors with factor 2)

General informations:

Controlled diaphragm pump

Hose length max. 7.8m (corresponds to two probe hose extensions)

Max positive pressure/flue gas +50mbar

Max negative pressure/flue gas -200mbar

User-defineable fuels 10 user-defineable fuels incl. test gas as fuel
Oper. temp.

-5 to +50°C

Accessories

Name
testo printer with wireless IRDA and infrared interface
Spare thermal paper for printer (6 rolls), permanent ink
SO2 cell 0-5000ppm (optional cell)
ISO calibration certificate for flue gas per analyzer box (up to 6 different cells with testo 350XL)
Pitot tube, length 350 mm (600°C)
Pitot tube, length 1000 mm (600°C)
Stainless steel pitot tube, length 350 mm (-40...1000°C)
Stainless steel pitot tube, length 1000 mm (-40...1000°C)
Connection hose, 5m silicon hose for pitot tubes
Spare sintered filters (2 off) for dusty flue
Adapter non-heated
Extension pipe, 1m long (1200°C), Inconel 625
Non heated sampling pipe, 1m long (600°C), stainless steel 1.4571
Non heated sampling pipe, 1m long (1200°C), Inconel 625
Non heated sampling pipe, 1m long (1800°C), Al-Oxide
Pre-filter for dusty flue gases, (1000°C), ceramic. Can only be mounted on extension pipe 0600 7802 and 0600 7804!!!
Thermocouple, -200° to 1000°C, 1.2 m long
Thermocouple, -200° to 1000°C, 2.2 m long
Thermocouple, -200° to 1000°C, 3.2 m long
Mounting flange, stainless steel 1.471, adjustable quick-action fitting suitable for all sampling/ extension pipes
Filters (10x) for filter trap (p/n 0554 0066)
CO cell, 0 to 10000 ppm
COlow cell, 0 to 500 ppm
NO cell, 0 to +3000 ppm NO
NOlow cell, 0 to +300 ppm NO

Technical Data

 

Meas. range
Accuracy
Resolution
Response time

O2 meas.

0 to 25Vol.%
±0.2Vol.%
0.01Vol.%
t90<20s

CO meas. (H2 compensated)

0 to 10000ppm

±10ppm or ±10% of mv (0 to 200ppm)

±20ppm or ±5% of mv (201 to 2000ppm)

±10% of mv (2001 to 1000ppm)

1 ppm

t90<40s

COlow meas. (H2 compensated)

0 to 500ppm

±2ppm (0 to 39.9ppm)

±5% of mv (2001 to 1000ppm)

0.1 ppm

t90<40s

NO meas.

0 to 3000ppm

±5ppm (0 to 99ppm)

±5% of mv (100 to 1999ppm)

±10% of mv (2000 to 3000ppm)

 

1 ppm

t90<30s

NOlow meas.

0 to 300ppm

±2ppm (0 to 39.9ppm)

±5% of mv (remaining range)

0.1 ppm

t90<30s

NO2 meas.

0 to 500ppm

±10ppm (0 to 199ppm)

±5% of mv (remaining range)

0.1 ppm

t90<40s

SO2 meas.

0 to 5000ppm

±10ppm (0 to 99ppm)

±5% of mv (remaining range)

 

1 ppm

t90<40s

Temp. meas. probe type

-40 to +1200°C

±0.5°C (0 to 99°C)

±0.5% of mv (remaining range)

0.1°C

 

Draught meas.

-40 to +40hPa

±0.03hPa (-2.99 to +2.99hPa)

±1.5% of mv (remaining range)

0.01 hPa

 

Differential pressure meas.

-200 to 200hPa

±0.05hPa (-49.9 to +49.9hPa)

±1.5% of mv (remaining range)

0.1 hPa

 

Absolute pressure meas.

600 to +1150hPa

 

±10 hPa

1 hPa

 

Derived parameters

 

Efficiency

Flue gas loss

Flue gas dewpoint

 

 

0 to 120%

0 to 99.9%

0 to 99.9°C

 

 

 

0.1%

0.1%

0.1°C

 

CO2 meas. (calculation from O2)

Response time t90=<40s

0 to Co2 max.

±0.2 Vol.%

0.1 Vol.%

 

PDF's


Click here for the pdf brochure Click here for more informations

Measure. range extension

Single dilution factor 5 (standard)

CO meas. (H2 compensated)

Meas. range

Accuracy

Resolution

700ppm to 50000ppm

±10% of mv (additional error)

1 ppm

COlow meas. (H2 compensated)

Meas. range

Accuracy

Resolution

500ppm to 2500ppm

±10% of mv (additional error)

0.1 ppm

NO2 meas.

Meas. range

Accuracy

Resolution

200ppm to 2500ppm

±10% of mv (additional error)

0.1 ppm

SO2 meas.

Meas. range

Accuracy

Resolution

500ppm to 25000ppm

±10% of mv (additional error)

1 ppm

Dilution of all sensors, Factor 2

(option 0440 3350)

O2 meas.

If meas. range extension is activated on all sensors: Accuracy:

±1Vol.% additional error (0 to 4.99Vol.%)

±0.5Vol.% additional error (5 to 25Vol.%)

CO meas. (H2 compensated)

Meas. range

Accuracy

Resolution

700ppm to 20000ppm

±10% of mv (additional error)

1 ppm

COlow meas. (H2 compensated)

Meas. range

Accuracy

Resolution

500ppm to 1000ppm

±10% of mv (additional error)

0.1 ppm

NO meas.

Meas. range

Accuracy

Resolution

500ppm to 6000ppm

±10% of mv (additional error)

1 ppm

NOlow meas.

Meas. range

Accuracy

Resolution

300ppm to 600ppm

±10% of mv (additional error)

0.1 ppm

NO2 meas.

Meas. range

Accuracy

Resolution

200ppm to 1000ppm

±10% of mv (additional error)

0.1 ppm

SO2 meas.

Meas. range

Accuracy

Resolution

500ppm to 1000ppm

±10% of mv (additional error)

1 ppm